The updated manual explicitly lists only FA-45-18-11, F-45-30-11, and F-45-48-11 rotors. The 5424 uses a different spindle taper. Forcing a 5424 rotor will permanently damage the 5402’s motor shaft.
Updated versions (including recent 2025/2026 revisions) focus on clarifying compliance with evolving CE safety directives (specifically IEC 61010-2-020) and updating maintenance procedures for older units.